Resetting the SpectroDens (from Techkon’s support page)

Do you want to reset your SpectroDens?

Resetting your SpectroDens will bring the device back to the factory default settings. **Please note that if you do this, all calibration data (including slope values), stored measurement values, reference values, target values, and tolerance values will be deleted.** If you have any questions at all, please get in touch with Techkon’s service team by calling 978-777-1854, chatting, or emailing service@techkonusa.com!

Some reasons you might want to reset your device are:

  1. Absolute white calibration error

  2. Measurements seem to be inaccurate

  3. Device is not recognized by your PC

Reset instructions:

  1. In the device settings menu, press the reset button.

  2. If the reset procedure in the settings window is not possible, a hardware reset has to be carried out.

  3. A hardware reset is done by pressing the red enter button and the green measurement button simultaneously for 8 seconds.